Low IOP is measured with great accuracy using non-contact applanation tonometer - Assessment uses corneal properties to determine pressure, may provide new gold standard - Modern Medicine
Low IOP is measured with great accuracy using non-contact applanation tonometerAssessment uses corneal properties to determine pressure, may provide new gold standard

Source: Ophthalmology Times


Key iconKey Points

  • Corneal stiffness forces vary over a wide range, causing both over- and under-estimation of the true IOP.
  • The biomechanical properties of the cornea, such as tear film and corneal resistance, can significantly affect the true IOP value, and using a device that appropriately and accurately compensates for this is clearly the optimal IOP measurement technique.
  • Every instrument that measures through the cornea will produce an IOP value that is affected by a combination of corneal and pressure properties.
  • The IOPcc measurement is independent of central corneal thickness.

Fort Lauderdale, FL—A recent study found a non-contact applanation tonometer (Ocular Response Analyzer [ORA], Reichert) to be more accurate in the assessment of IOP when compared with the Goldmann applanation tonometer (GAT), particularly when measuring for extremely low and negative IOP, according David A. Luce, PhD, speaking at the annual meeting of the Association for Research in Vision in Ophthalmology.


Dr. Luce
Long considered to be the gold standard technique in measuring IOP, the GAT was designed to achieve a cancellation or correction of tear film and cornea resistance forces for an average cornea in order to achieve accurate IOP measurement results. Most of the currently available tonometers are calibrated by matching GAT measurements in a clinical setting.

"Corneal stiffness forces vary over a wide range, causing both over- and under-estimation of the true IOP," said Dr. Luce, chief scientist, Reichert. "The significant impact of the corneal biomechanical effect on the measurement of extremely low and negative IOP has become more apparent to ophthalmologists, and this realization has led to the dawn of other IOP assessment devices such as [this non-contact applanation tonometer], which can achieve more accurate IOP values by determining the negative Goldmann-equivalent IOP (IOPg) and more accurately compensate for corneal resistance."

Extremely low IOP

In a recent study, Dr. Luce and colleagues evaluated the effectiveness of the non-contact applanation tonometer in measuring extremely low IOP in eyes with "soft" corneas. Twelve eyes (nine with primary open-angle glaucoma, one with dry eye, and two normal eyes) with negative IOPg (defined as less than 1 mm Hg) were selected for analysis in order to determine the effect of corneal biomechanical compensation using this non-contact applanation tonometer's IOPg and corneal-compensated IOP (IOPcc) parameters. Also, the progression over time of the IOPg and IOPcc was evaluated for two of the participants.

Results for the 12 eyes showed a mean IOPg of –1.40 mm Hg with a standard deviation of 1.29 mm Hg, ranging from –4.98 to –0.07 mm Hg. The mean IOPcc was +3.02 with a standard deviation 1.66 mm Hg, ranging from +0.2 to +6.3 mm Hg. The IOPcc was greater than zero in all of the eyes, and for the two patients who underwent multiple measurement sessions following glaucoma surgery, the IOPcc showed a regular increasing "recovery" IOP.

The IOP measurements taken by ophthalmologists are generally accepted to be the true values; however, because very low and negative IOP values can occur when assessing IOP, why they occur is of particular interest. According to Dr. Luce, very low and even negative IOP readings are not at all uncommon following a trabeculectomy. Whatever the cause, the biomechanical properties of the cornea, such as tear film and corneal resistance, can significantly affect the true IOP value, and using a device that appropriately and accurately compensates for this is clearly the optimal IOP measurement technique.

"Even though extremely low and negative IOP values are an unusual situation, the fact remains that they do occur and can be due to these properties of the cornea," Dr. Luce said. "After we understand why this happens, it would be advantageous to use a device that can accurately compensate for the corneal effect and subsequently produce a true IOP value."


post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All comments submitted are subject to review, and may be delayed before posting. We reserve the right not to post comments.

eSamples check closet
eSamples check closet
Practice ToolsPractice Tools
Coding Counselor
Coding Counselor

Simple and accurate ICD-9 code search. Start Here

Formulary Counselor
Formulary Counselor

Find health plan drug coverage in your area. Start Here

Patient Education
Patient Education

Print customized patient education handouts. Start Here

Surgical Video Center
Surgical Video Center

On-demand surgery demos and presentations. Start Here



Source: Ophthalmology Times,
Click here